Lightning Surge Simulator (LSS) Lightning Surge Simulator (RINGWAVE generator equipped)

Features
- Tester compliant with IEC 61000-4-5 Ed.3 standard
- The tester can generate 0.5μs-100kHz RINGWAVE as specified in IEC61000-4-12 standard.
(Maximum voltage surge: 12kV high voltage test can be performed.) - Maximum output voltage: 15kV: In reliability evaluation tests for lightning surge tests, destructive tests can also be included.
- Adoption of a large LCD screen operation panel: A large LCD panel screen is used in the operation section for improved visibility and operability.
- MPU control for easy continuous testing: Surge output, waveform switching, polarity switching, and sequential operation can be performed automatically.
- Manual and program modes are available: Manual mode for standardized tests and single-shot tests, and program mode for continuous tests under different conditions. Test conditions can be easily set to suit the application.
- Excellent safety features such as interlock
- Waveform check terminal provided as standard package
- Output waveforms can be checked using your oscilloscope and a BNC cable.
- Isolation transformers for EUT power supply protection are available. (Optional)
*This product is a custom specification product. Please contact us for details.
Specifications
■Surge Generating Part
PARAMETER SPECIFICATIONS REMARKS 1.2/50μs -8/20μs
Combination waveformsOpen voltage 0.5kV~15kV ±10% Cable length:One side 0.5m
Voltage step: 0.1kV
Setting available from 0kVFront time 1.2μs±30% Time to half-value 50μs±20% Short-circuited current 250A~7500A ±10% Front time 8μs±20% Time to half-value 20μs±20% 0.5us-100kHz
RING WAVEOpen voltage 0.5kV~12kV ±10% Cable length:One side 0.5m
Voltage step: 0.1kV
Setting available from 0kVRise time 0.5μs±30% Ringing frequency 100kHz±10% Waveform envelope 2nd peak:40% to 110% of the 1st peak
3rd peak:40% to 80% of the 2nd peak
4th peak:40% to 80% of the 3rd peakShort-circuited current (Set voltage/12Ω)±10%
(Set voltage/30Ω)±10%Short-circuited current Front time 0.2μs~1μs Output polarity Positive / Negative Output impedance 2Ω±10% 1.2/50μs waveform Generating circuit Floating Minimum charging
period0.0kV ~6.0kV:10sec
6.1kV ~15.0kV:20sec1.2/50μs waveform 0.0kV ~6.0kV:1sec
6.1kV ~12.0kV:5secRING WAVE ■AC / DC Line Injection Part
PARAMETER SPECIFICATIONS REMARKS 1.2/50μs -8/20μs
Combination waveformsOpen voltage 0.5kV~15kV ±10% Coupling circuit:18μF
Decoupling coil:1.5mH
Cable length:One side 0.5m
Setting available from 0kV
Line input ports openFront time 1.2μs±30% Time to half-value 50μs+10μs /-10μs Short-circuited current 250A~7500A ±10% Front time 8μs±20% Time to half-value 20μs±20% Open voltage 0.5kV~15kV ±10% Coupling circuit:10Ω+9μF
Decoupling coil:1.5mH
Cable length:One side 0.5m
Setting available from 0kV
Line input ports openFront time 1.2μs±30% Time to half-value 50μs+10μs /-25μs Short-circuited current 41.7A~1250A ±10% Front time 2.5μs±30% Time to half-value 25μs±30% 0.5us-100kHz
RING WAVEOpen voltage 0.5kV~12kV ±10% Coupling circuit:4.5μF
Decoupling coil:1.5mH
Cable length:One side 0.5m
Setting available from 0kV
Line input ports openRise time 0.5μs±30% Ringing frequency 100kHz±10% Waveform envelope 2nd peak:40% to 110% of the 1st peak
3rd peak:40% to 80% of the 2nd peak
4th peak:40% to 80% of the 3rd peakShort-circuited current (Set voltage/12Ω)±10%
(Set voltage/30Ω)±10%Short-circuited current Front time 0.2μs~1μs Injection mode Line – line 1.2/50μs waveform
Coupling circuit:18μF
(10Ω+9μF selectable)RINGWAVE
Coupling circuit:4.5μFLine - PE 1.2/50μs waveform
Coupling circuit:10Ω+9μF(18μF selectable)
L/N injected simultaneously:
each line 9μF(up to 8kV)RINGWAVE
Coupling circuit:4.5μF
L/N injected simultaneously:
each line 4.5μFNumber of Phases Single Phase LINE configuration AC: L/N/PE
DC: +/-/PEPower capacity for EUT line AC240V/30A MAX 50/60Hz
DC125V/20A MAXDecoupling coil 1.5mH Residual voltage Less than 15% of the injected voltage or less than double of the rated voltage (peak value) Phase angle control 0°~ 360°±10° EUT power supply: More than AC90V
Working at 50Hz/60Hz±10%■General Specifications
GENERAL SPECIFICATIONS Driving power
supplyAC100V~AC240V ±10%
50Hz / 60HzConsumption power 400VA Operational environment Temperature:15~35℃
Humidity:25~75%RHDimensions W555×H1800×D790 mm
Protrusions excludedMass Approx. 300kg SAFETY FUNCTIONS Emergency stop Push-lock type switch: Stops the test, turns OFF the high voltage generating part, and cuts OFF EUT line. EUT line ON / OFF: selectable Interlock function Detects the surge output connector status, the protective door status, and the status of connectors for connecting external equipment. Warning lamp LED start blinking when a test starts Color: Red Alarm connector Available to connect the external warning lamp. The warning lamp starts illuminating when a test starts. OUTPUT MONITOR Voltage monitor BNC output: 2000V/V
Accuracy:±10% vs. actual output for 1.2/50 uS wave
Accuracy:±15% vs. actual output for ringwaveSurge out setting *
When output is open
Waveform not guaranteedCurrent monitor BNC output: 1000A/V
Accuracy:±10% vs. actual output
Accuracy:±15% vs. actual output for ringwaveSurge out setting *
When output is short-circuited
Waveform not guaranteed
* Only the peak value is guaranteed when the output is open or shorted. Performance is not guaranteed when an EUT is connected.Contact
Repair, Inspection and Calibration
Customer Service CenterNew Purchase Consultation, Materials Request
Product Inquiry Form
Options
-
For Lightning Surge Simulator CDN UNIT (AC600V / 300A)
This equipment can be combined with the LSS-F03 series lightning surge simulators to perform noise superimposition tests on three-phase four-wire (L1, L2, L3, N) lines up to AC600V / 300A.
* This product is a custom product. Please contact us for details. -
CDN dedicated to DC power supply coupling for lightning surge tester (DC500V / 60A)
This tester by combining the lightning surge tester, it is possible to perform noise superim-position tests up to DC500V 60A.
In combination with the lightning surge tester body, you can perform tests by switching the applied phase and setting the applied phase sweep.
※When using this superimposition unit, it is necessary to modify the tester body. Please con-tact us for details. -
Telecom waveform check cable set MODEL : 05-00150A
Jig used for measureing the output waveform from the CDN for telecom line. -
Waveform Checking Cables Set MODEL : 05-00099A
Jig for checking voltage waveforms and current waveforms of LSS-F03 series. -
Terminal Connection Board with Multi-Outlet(3P) MODEL : 18-00048B
A relay terminal board for connecting the output of the LSS-F03 series to the EUT.
By wiring to the included multi-outlet, you can directly connect a power plug that supports the standards of each country.
single phase 3 lines (withstand voltage 4.5kV)
● Compatible models : LSS-F03 series -
Terminal Block for 3P MODEL:18-00047A
Terminal block board for CDN to connect EUT. 3 pins
● Compatible models : LSS-F03 series -
Terminal Block for 5P MODEL:18-00044A
Terminal block board for CDN to connect EUT. 5 pins
● Compatible models : LSS-F03 series -
DC line input cable MODEL : 05-00136A
-
AC line input cable (3-phase) MODEL : 05-00135A
-
AC Line Input Cable (Single phase) MODEL : 05-00134A
-
Arrester capacitor unit MODEL : 08-00016A
Arrester unit for surge coupling
● Compatible models : LSS-F03 series -
Arrester capacitor unit MODEL : 08-00012A
Arrester unit for surge decoupling.
● Compatible models : LSS-F03 series -
USB Optical Module Kit MODEL : 07-00022A
Connection adapter used for remotely controlling the simulator from a PC.
Equipped with USB-Optical conversion fiber optic cable (5m).
● Compatible models : LSS-F03 series -
Warning Lamp MODEL : 11-00017A
Alarm lamp for LSS series. Allows to alert and call for attention by blinking during the test.
● Compatible models : LSS-F03 series -
Protective Safety Fence MODEL : 11-00010A
Allows construction of a safe test environment by connecting with the lightning surge simulator's interlock function.
Combined use with the EUT protection box ensures a completely safe test environment. -
EUT Protective Safety Box MODEL:11-00006A
Protection box to prevent access to EUT during the test.
Further safety can be achieved by combining with the
protective safety is fence
(W) 600 x (D) 400 x (H) 350mm *protrusions excluded -
CDN for Interconnection Lines MODEL : LSS-INJ6401SIG
This CDN product is used to apply surges to interconnection lines as defined in the IEC61000-4-5 Standard. With The EUT power capacity of DC50V / 1A it is possible to inject surges to interconnection lines up to 6,600V. Possible to bypass inductor (20 mH) with connecting the attached connection plug to inductor bypass terminal in DC output. Possible to equip the attached surge protective arrestor between each line and ground. -
Isolation Transformer MODEL : TF-6633P
Model TF-6633P is three-phase isolation transformer rated AC 600 V / 63 A and dielectric strength of 4 kV. For safety reasons, an isolation transformer is indispensable for AC powered testing for equipment. -
Isolation Transformer MODEL : TF-6503P
Model TF-6503P is three-phase isolation transformer rated AC 600 V / 50 A and dielectric strength of 4 kV. For safety reasons, an isolation transformer is indispensable for AC powered testing for equipment. -
Isolation Transformer MODEL : TF-2302P
Model TF-2302P is a single-phase isolation transformer rated AC240V/30A with dielectric strength of 4kV.
For safety reasons, an isolation transformer is indispensable for AC powered testing for equipment. -
High-speed communication lines CDN MODEL : F-130814-1004
This CDN product is used to apply surges to unshielded symmetrical high-speed communication lines with speed up to 1000Mbit/s, as defined in the IEC 61000-4-5 Standard.
-
AC line coupling unit for LSS-700 series (Three-phase AC600V / 50A)
This tester can perform a noise coupling test on a three-phase line up to AC600V / 50A by combining with the lightning surge tester (LSS-720B).
※This is a customized product. Please contact us for details. -
External trigger box for LSS-700 series
This is a box that can turn on the trigger (generate a surge) at a specified number of times and at specified intervals by combining it with the lightning surge tester LSS-700 series.
Related Products
-
Lightning Surge Simulator (AC500V/200A DC600V/200A)
read more -
This simulator simulates "high-energy induced lightning noise" induced in distribution lines and communication lines due to ground potential fluctuations caused by lightning strikes, and evaluates the resistance of electronic devices.
It is possible to check the dielectric strength due to induced lightning at a level that cannot be confirmed with the combination waveform required by the IEC standard.Lightning Surge Simulator LSS-720B2
read more -
For a stricter test with a maximum voltage of 15 kV
A tester simulatively generates "High energy induced lightning noise" which induced to distribution lines or communication lines by ground potential fluctuation caused by lightning strikes.Lightning Surge Simulator LSS-F03 series
read more
News
- 02/04/2025 Information
Join Telonic/NoiseKen at Southern Manufacturing & Electronics 2025!
- 12/04/2024 Information
Information on year-end and New Year holidays
- 10/29/2024 Products & Services
5GHz to 8GHz Broadband sleeve antenna now on sale
- 10/02/2024 Exhibitions & Events
Optimizing PCB Design: From Schematic to Prototype Testing with NoiseKen EPS and ZUKEN CR-8000
Contact
Repair, Inspection and Calibration
Customer Service CenterNew Purchase Consultation, Materials Request
Product Inquiry Form
-
ESD Simulator ESS-S3011A/GT-30RA
https://www.youtube.com/watch?v=kdKCHXaVhf8